Chapter 3: Objective Maintenance
5063669 E 13
Installing an Objective in an Empty Slot
You must install objectives in specific slots in the turret. See Installing an Objective on page
12 for details.
To install an objective in an empty slot:
1.
In the CellReporterXpress software, on the Home page, click Devices.
2.
Click Show Device Options to expand the details for the instrument where you
want to install an objective.
3. Click the Objectives tab.
4.
In the tile for the objective slot where you want to install, click Component
Exchange.
5. Click the Choose Objective drop-down list box, and select the objective you want to
install.
6. Click Open Maintenance Door.
7. Install the objective in the slot by gently turning it clockwise.
CAUTION! Retain the objective case for future storage needs. When not installed
in the instrument, an objective should always be stored in its case.
8. Do both of the following:
a. Manually close the maintenance door.
b. In the CellReporterXpress software, click Close Maintenance Door.
9. Click Apply.
After you install a new objective from an after-sales purchase, you must calibrate it. See
Calibrating an Objective on page 16 for details.

Other manuals for Molecular Devices ImageXpress Pico

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Molecular Devices ImageXpress Pico and is the answer not in the manual?

Molecular Devices ImageXpress Pico Specifications

General IconGeneral
BrandMolecular Devices
ModelImageXpress Pico
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals