Chapter 3: Objective Maintenance
5063669 E 17
4. Note the status of the objective, which is indicated in the tile.
If the tile indicates Not Calibrated, continue to the next step to perform the
calibration procedure.
If the tile indicates Calibrated, you typically do not need to perform the calibration
procedure and you can skip this procedure.
Note: An exception to this is when you replace an objective with another
objective of the same magnification. In this case, you should continue to the next
step and perform the calibration procedure regardless of the label in the tile.
5.
In the tile for the objective you want to calibrate, click Objective Calibration.
6. Follow the on-screen instructions to complete the calibration.
Tip:
In the Choose Slide Holder drop-down list box, select 4 Slide Holder.
Insert the slides face down in the slide holder (that is, printed side down).
When each step completes, click Next to continue to the next step.

Other manuals for Molecular Devices ImageXpress Pico

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Molecular Devices ImageXpress Pico and is the answer not in the manual?

Molecular Devices ImageXpress Pico Specifications

General IconGeneral
BrandMolecular Devices
ModelImageXpress Pico
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals