EasyManuals Logo

Morita veraviewepocs 2d User Manual

Morita veraviewepocs 2d
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
Panorama Exposures
20
Do not bump, jiggle, vibrate or use excessive force.
Use both hands to put the cassette in and take it out.
The cassette weights about 2 kg, and it could injure
your foot if you dropped it. The sensor would also be
damaged.
Do not swing the cassette around. The vibrations
could disturb the sensors adjustment.
Do not store under conditions of high temperature and
humidity.
Storage Temperature; -10°C to +50°C (+14°F to +122°F)
Storage humidity: 5% to 85% relative humidity with no
condensation.
No frequent or continuous exposure to direct sunlight.
Do not get water, detergents or chemical solutions on
the cassette. These could seep inside and damage it.
Put the cassette in and take it out with great care.
Using excessive force could damage the connection
components.
Clean by wiping with a soft, dry cloth. Do not touch the
connection components directly with bare fingers etc.
Static electricity could damage the sensor.
Do not allow dust or other debris to collect on the
connection components; this could cause poor
contact.
Slide the cassette all the way into the holder and make
sure it is locked into place.
Otherwise images could be distorted, transmission
could be interrupted, or equipment could be damaged.