Panorama Exposures
27
Take care that the temple stabilizers do not strike the
patient in the eye.
Do not use excessive force to close the temple
stabilizers.
This could be uncomfortable for the patient or damage
the stabilizers.
Forcing the patient in or out could also damage the
stabilizers.
3. Use the knob to open the temple stabilizers. Have the
patient move forward without slouching or otherwise
changing his posture. Then have him put his chin on the
chin rest and lightly grip the patient handles. Make sure his
shoulders are lowered.
4. Close the temple stabilizers until they lightly touch the
patient’s temples.
5. Make sure the patient’s face is straight and then line up the
Frankfort plane beam with the patient’s frankfort plane
using the beam Up or Down key.
CLASS 2 LASER PRODUCT: A class 2 laser is used for
the positioning beams.
The laser beams could damage the eyes.
Do not stare into the positioning beams. Warn the patient
not to look at the positioning beams.
Chin Rest
Patient Handle
Temple Stabilizer Knob
Frankfort Plane Beam
Temple Stabilizer