Panorama Exposures
50
Open Mouth Exposure (Second Exposure)
1. Have the patient open their mouth.
2. Pick up the handswitch and hold down the emission button.
The arm will start to rotate and exposures will be made of
the left and right sides. Release the emission button when
the arm stops rotating.
The Ready LED on the control box will go out to show that
the exposure procedure has been completed.
Always leave the X-ray booth, and press the emission
button outside of it.
In case of an emergency, release the emission button;
this will completely stop the unit.
3. Open the temple stabilizers all the way up and guide the
patient away from the unit.
Press the Ready key on the patient frame or the arm
operation panel.
The arm will go back to the patient entry position.
4. Remove the lip-nose rest and replace it with the chin rest.
Emission LED
Emission Button