Computer Screen and Arm Display
Problem Possible Cause Remedy
In computer screen for CT images
(for CT exposure)
In X-ray unit Arm Display
The computer is not receiving the
image. X-ray is terminated.
LAN cable
transmission problem
* LAN cable is not properly
connected.
Reconnect and confirm LAN
cable connection.
Restart i-Dixel application.
If either of the following error messages
appear in the i-Dixel computer screen
* If the above error messages appear the
following message may appear in the
Arm Display:
Computer is not ready to capture
images.
This happens when the
computer runs out of memory
due to repeated exposures and
image reconstruction.
Restart i-Dixel application