Message Numbers
Messages appear in the Arm Display when anomalies are detected. Respond according to the error message
number as explained in the following chart. If this does not solve the problem, contact your local dealer or J.
MORITA OFFICE. Make a note of the error number and report it when requesting help for the company.
Possible Cause Remedy
00
Overcurrent protection for the low-voltage circuit
is activated.
Turn off, wait 1 minute, turn back on
01
Short pin is not installed on DCN12 of the
CPU1board.
Contact your local dealer or J. MORITA
OFFICE.
02
X-ray head overheat! Leave the unit at least 30
minutes to cool down.
Wait at least 30 minutes for the unit to cool
down. Overheating is due to excessively
frequent X-ray emission. This will
eventually degrade the tube performance
and should be avoided.
03
X Axis of the Arm did not return to its original
position in specified time.
Press Ready key and put unit into Ready
mode
04
Y Axis of the Arm did not return to its original
position in specified time.
Press Ready key and put unit into Ready
mode
05
Arm did not return to its original position in
specified time.
Press Ready key and put unit into Ready
mode
07
AF Beam did not return to its original positioning
specified time.
Press Ready key and put unit into Ready
mode
08
Horizontal slit did not return to its original position
in specified time.
Turn off, wait 1 minute, turn back on
09
X-ray Head did not return to its original angle in
specified time.
Turn off, wait 5 seconds, turn back on
C1
No communication with the host computer.
Error C1 occurs when there is no response
from the application software. Make sure
that the computer and the HUB are turned
on and that the application is running.
Then turn the X-ray unit off and back on
again.
C2
Transmission speed of the HUB doesn't match.
Remove the cassette, press Ready key to
reset the error, and then reinsert the
cassette. Check connections between
HUB and main X-ray unit.
C3
No response from HUB.
Check connections between HUB and
main X-ray unit.