91
3. Verification Procedure
1) Compare the DAP values from the previous two scans; scan with the copper plate (1) and with the copper
plate (1) + (2).
2) Check if the value with the copper plate (1) + (2) is greater than the one with the copper plate (1).
3) Use the following flowchart to verify the DDAE.
START
NO
(1) + (2) > (1)
YES
PASS FAIL
4) Results
PASS : DDAE operates correctly.
FAIL : Try the same procedure carefully again.
If still not improved, please contact your local sales representative.