TMJ Quadruple (2/4 images):
Patient size
Super high resolution
Mode
4.3s 8.6s
Cephalometric:
Accuracy of displayed values: ± (5 % + 50 ms) (* Registered value for FDA is ± 10 %)
Test instruction of X-ray tube voltage, current, and exposure time Constant (manual) exposure mode
Reproducibility of air karma Coefficient of variation max. 0.05
Minimum mAs Dental Arch Panoramic: 6.0 mAs
Cephalometric: 3.5 mAs
Emission Button Dead Man Type
Auto Exposure (Not applied for Cephalometric)
Setting parameters +4 to -4 (9 steps) Digital Display
Maximum possible excursion 60 to 80 kV
1 to 10 mA
Reproducibility of air karma Coefficient of variation max. 0.05
Minimum exposure time 4.3 seconds (Exposure time is fixed value, not auto exposure factor.)
Verification method by test piece