41
Settings Parameters
GB
Model 9010/9020 SIL
5.6 Access Code 7 - Peripheral Addresses
Step Function Note
P1
RS485 Address
Permissible values 1-247
P2
RS485 Baud Rate
Permissible values 9600 or 38400 Bd
Default setting is 38400
P3
Ethernet Configuration
OFF - Ethernet is not enabled
dhcP - IP Address provided by DHCP server
Stc - Use static IP address
Default: DHCP
P4
Static IP Address
setting the address in "aaa/bbb/ccc/ddd" format
P5
Subnet Mask
setting the address in "aaa/bbb/ccc/ddd" format
P6
Gateway IP Address
setting the address in "aaa/bbb/ccc/ddd" format
P7
DNS IP Address
setting the address in "aaa/bbb/ccc/ddd" format
P8
Show IP address
IP address is displayed step by step in "aaa/bbb/ccc/ddd" format