MSA
Approvals
PrimaX
52
GB
7.2 Marking and Certificates according to IECEx
PrimaX P
Manufacturer: MSA AUER GmbH
Thiemannstraße 1
D-12059 Berlin
Product: PrimaX P
Type of protection: IEC 60079-0:2007, IEC 60079-1:2007,
IEC 60079-11:2006, IEC 60079-31:2009
Measuring function for explosion protection: no
Gas: see manual
Marking: Prima X P
Ex d [ia] IIC T4 Gb
Ex t [ia] IIIC T130°C Db
-40°C Ta +70°C
P
i
770 mW, U
i
28 V, I
i
110 mA, L
i
= 0, C
i
= 0
Option: HART Module Connector, only for temporary connection of
an intrinsic safe HART Field Communicator
P
o
200 mW, U
o
2,7 V, I
0
140 mA, L
o
10 µH, C
o
1nF
P
i
5mW, U
i
5 V, I
i
1 mA, L
i
= 0, C
i
= 0
IEC-Type Examination Certificate:IECEx BVS 10.0043 X
Quality Assurance Notification: 0158
Year of Manufacture: see Label
Serial Nr.: see Label