46
4 Operation US
Gas
(Code)
Range Range
Default
Unit
Default
Warn
Default
Warn
Default AO
1
(mA)
Alarm
Default
Alarm
DefaultAO
1
(mA)
Alarm
Action
Default
Cal
Level
Default
Alarm
Min
Alarm
Max
Cal
Level
Min
Cal
Level
Max
Display
Resolution
2
Dioxide
(R51)
IR700
Carbon
Dioxide
(R52)
0-5000 0-5000 PPM 1500 8.8 3000 13.6 Increasing
Non-Latch
2500 250 4750 Fixed at 2500
PPM
50
IR700
Carbon
Dioxide
(R53)
0-
10000
0-10000 PPM 3000 8.8 6000 13.6 Increasing
Non-Latch
5000 500 9500 Fixed at 5000
PPM
100
IR700
Carbon
Dioxide
(R54)
0-
30000
0-30000 PPM 9000 8.8 18000 13.6 Increasing
Non-Latch
15000 1500 28500 Fixed at 15000
PPM
300
IR700
Carbon
Dioxide
(R55)
0-
50000
0-50000 PPM 15000 8.8 30000 13.6 Increasing
Non-Latch
25000 2500 47500 Fixed at 25000
PPM
500
1
- Analog Output for default range and alarms
2
- Display resolution is not a configurable option