MSA S5000 - Table 14 Passive Sensor Default Settings

MSA S5000
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
46
4 Operation US
Gas
(Code)
Range Range
Default
Unit
Default
Warn
Default
Warn
Default AO
1
(mA)
Alarm
Default
Alarm
DefaultAO
1
(mA)
Alarm
Action
Default
Cal
Level
Default
Alarm
Min
Alarm
Max
Cal
Level
Min
Cal
Level
Max
Display
Resolution
2
Dioxide
(R51)
IR700
Carbon
Dioxide
(R52)
0-5000 0-5000 PPM 1500 8.8 3000 13.6 Increasing
Non-Latch
2500 250 4750 Fixed at 2500
PPM
50
IR700
Carbon
Dioxide
(R53)
0-
10000
0-10000 PPM 3000 8.8 6000 13.6 Increasing
Non-Latch
5000 500 9500 Fixed at 5000
PPM
100
IR700
Carbon
Dioxide
(R54)
0-
30000
0-30000 PPM 9000 8.8 18000 13.6 Increasing
Non-Latch
15000 1500 28500 Fixed at 15000
PPM
300
IR700
Carbon
Dioxide
(R55)
0-
50000
0-50000 PPM 15000 8.8 30000 13.6 Increasing
Non-Latch
25000 2500 47500 Fixed at 25000
PPM
500
1
- Analog Output for default range and alarms
2
- Display resolution is not a configurable option