4.2.12 Transmitter Setting Reset
The transmitter default values can be restored by resetting the main board. The unit will restart after the reset and the unit may
go into sensor fault. During transmitter reset, the analog output displays 1.25mA.
NOTICE
Verify all sensor settings (calibration level and alarm values) after resetting transmitter. Sensors may require recalibration to
clear a fault after resetting transmitter.
4.3 Info Menu - Viewing Device Status
This section describes how to view the following options:
• Display Current System Time
• Change System Time
• Sensor Life and Health Status
• Last Calibration Dates
• Non-critical Faults
• Device Tag
• Bluetooth ID
4.3.1 AO Type
1. Go to Info menu and select AO Type.
The AO type selected for this unit, Source or Sink, will scroll across the screen twice.
2. Touch button when "Finished?" is displayed to exit.
4.3.2 View or Change System Time
1. Go to Info menu and select System Time.
2. Hold finger on button while menu selections scroll across screen.
3. Remove finger when Info is displayed.
4. Touch button when System Time is displayed.
The current date followed by the current time will scroll across the screen twice.
5. Touch button when Change Time displays if the current date and time are not correct.
6. If selected, change the following values using the button and selecting "Finished?" after each entry:
a. Month
b. Day
c. Year
d. Hour
e. Minutes
7. Touch button when "Finished?" is displayed to exit.
You can also use the X/S connect app to sync time and date with a mobile device.