Sensors & Patient Cables (Pulse Oximeter)
These sensors are recommended because they are compatible with the
Dolphin CPAP and are often appropriate for the Dolphin CPAP’s target patient
population. This is NOT a complete list of sensors compatible with the Dolphin
CPAP. Contact your MTTS representative for a full list of compatible sensors.
Reusable sensors can be cleaned per the following procedure:
Remove the sensor from the patient.
Disconnect the sensor from the Dolphin CPAP.
Allow the sensor to air dry before returning it to operation.
Single use sensors may be reapplied to the same patient if the emitter and
detector windows are clear and the adhesive still adheres to the skin.
If the sensor fails to track the pulse consistently, the sensors may be
incorrectly positioned. Reposition the sensor or choose a different monitoring
site.
To avoid cross contamination only use Masimo single use sensors on the same
patient.
Do not reprocess any single use sensors.
Do not soak or immerse the sensor in any liquid solution. Do not sterilize any
Masimo sensor by irradiation, steam, or ethylene oxide.
Cleaning and reuse of Masimo sensors
Reattachment of single use adhesive sensors
Sensors & Patient Cables (Pulse Oximeter)
Reusable patient cables of various lengths are available. Only use appropriate
Masimo oximetry cables for SpO
2
measurements. Other patient cables may cause
improper Pulse Oximeter performance.
RECOMMENDED PATIENT CABLES
Cable Description
M-LNC
TM
M-LNCS
TM
Series Patient Cable
This cable is recommended because it is compatible with the recommended
sensors. This is NOT a complete list of cables compatible with the Dolphin
CPAP. Contact your MTTS representative for a full list of compatible cables.
Patient cables can be cleaned per the following procedure:
Remove the cable from the sensor.
Disconnect the sensor from the Dolphin CPAP.
Allow the cable to dry before returning it to operation.
Carefully route patient cables to reduce the possibility of patient
entanglement or strangulation.
Do not soak or immerse patient cables in any liquid solution. Do not sterilize
patient cables by irradiation, steam, or ethylene oxide. See the cleaning
instructions in the Directions for Use for reusable Masimo patient cables.
Do not reprocess any Masimo SET patient cables.
Masimo SET patient cables
Cleaning and reuse of Masimo SET patient cables