Dolphin CPAP. User manual
46
Maintenance (CPAP)
The Dolphin CPAP includes a number of parts that must be replaced or serviced
during the lifetime of the device. These parts include:
Component Approx. Duration of Use Maintenance Information

2 years or 150 uses
Replace parts when need-
ed
PEEP chamber
Breathing circuit
Cleaning brush
 6 months
Clean/replace according
to step 1 in this section.
Oxygen sensor 18 months
Calibrate every 3 months
according to step 2 in this
section.
47
Maintenance (CPAP)
1. Air lter replacement
Turn off the device and disconnect the power cord.A.
B.
Use a Philips head screwdriver to
remove the screws on the bottom hatch
of the device. Remove the bottom hatch
completely and place it on the table
C.
    
and replace with a clean one.

