Chapter 4 Theory of Operation
SCXI-1120/D User Manual 4-4 www.natinst.com
Digital Control Circuitry
Figure 4-3 diagrams the SCXI-1120/D digital control.
Figure 4-3.
SCXI-1120/D Digital Control
The digital control section consists of the Configuration Register and the
Module ID Register.
The Configuration Register provides channel selection and configures the
SCXI-1120/D for scanning options. All the control bits are fed into a latch
before being routed to the rest of the module. The channel-select bits are
taken directly from the shift register.
The Module ID Register contains the module ID 4 hex, a code unique to the
SCXI-1120, or 44 hex, a code unique to the SCXI-1120D. You can read this
module ID over the SCXIbus to determine the type of module in a
particular slot.
Module ID Register
Configuration
Register
Serial Data Out
Input Channel
Select
Output
Stage
Control
Hardware
Scan
Control
SCANCLK
Path
Control
Buffered
Serial Data In
Buffered Digital
Control Signals

Table of Contents

Other manuals for National Instruments SCXI-1000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments SCXI-1000 and is the answer not in the manual?

National Instruments SCXI-1000 Specifications

General IconGeneral
BrandNational Instruments
ModelSCXI-1000
CategoryControl Unit
LanguageEnglish

Related product manuals