Chapter 4 Theory of Operation
SCXI-1120/D User Manual 4-8 www.natinst.com
The isolated amplifiers fulfill two purposes on the SCXI-1120/D module.
They convert a small signal riding on a high common-mode voltage into a
single-ended signal with respect to the SCXI-1120/D chassis ground. With
this conversion, the input analog signal can be extracted from a high
common-mode voltage or noise before being sampled and converted by the
DAQ device. The isolated amplifier also amplifies and conditions an input
signal, which results in an increase in measurement resolution and
accuracy.
After isolation, further filtering is available to increase the noise immunity
of the amplifier channel. It is important to note that the overall amplifier
bandwidth is determined by both filtering stages, so to achieve the required
bandwidth, both filtering sections should be set the same, as indicated in
Chapter 2, Configuration and Installation.

Table of Contents

Other manuals for National Instruments SCXI-1000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments SCXI-1000 and is the answer not in the manual?

National Instruments SCXI-1000 Specifications

General IconGeneral
BrandNational Instruments
ModelSCXI-1000
CategoryControl Unit
LanguageEnglish

Related product manuals