16 | ni.com | NI VirtualBench Calibration Procedure
Note Use the values in the 24 Hour Limits column for a post-adjustment 
verification only. Otherwise, use the values in the 1 Year Limits column.
7. Repeat steps 4 through 6 for each verification point listed in Table 8.
8. Reset the calibrator for safety reasons.
9. Disconnect all cables from the calibrator and VirtualBench.
Note Consult the user documentation for the amplifier used in step 10 for the 
recommended warm up time (at least 30 minutes) to ensure that the amplifier 
circuitry is at a stable operating temperature.
10. To verify DC current ranges ≥1 A, connect an amplifier to the calibrator, connect the 
connectors on one end of the low EMF copper cable to the VirtualBench A and LO banana 
plug connectors, and connect the connectors on the other end of the cable to the HI and LO 
binding posts on the amplifier. Figure 5 shows the correct connections.
Figure 5.  Cable Connections for DC Current Ranges ≥1A
11. Wait 2 minutes for the thermal EMF to stabilize.
12. Generate 1.01 A verification point on the calibrator.
13. Call 
DMM Configure Measurement with the following parameters:
• Function = 
DC Current
• Auto Range = FALSE
• Manual Range = 1
1 VirtualBench DMM
2 Multifunction Calibrator
3 Low EMF Copper Cable
4 Amplifier
5 Shielded Interface Cable
HI
LO
HI
LO
HVI
SENSE
VΩ
OUTPUT
VΩA
AUX
CURRENT
GUARD GROUND
HI
LO
5
1
2
4
3