58 | ni.com | NI VirtualBench Calibration Procedure
Note Use the values in the 1 Year Limits column. The values in the 24 Hour Limits
column are for a post-adjustment verification only.
8. Call
PS Read Output with the following parameter:
• Channel =
ps/+6V
9. Subtract the DMM reading from step 7 from the DC power supply Actual Voltage Level
reading in step 8. Verify that calculation falls between the limits in Table 20 or 21.
10. Repeat steps 5 through 9 for each verification point on this channel.
11. Call
PS Reset Instrument.
Table 21. VB-8034/8054 DC Power Supply Voltage Programming
and Measurement Accuracy
Channel
Verification
Point (V)
Current
Limit (A)
1 Year Limits ±(V)
24 Hour Limits
±(V)
+6 V 0 3 0.0050 0.0020
1.5 0.0065 0.0024
3 0.0080 0.0028
4.5 0.0095 0.0031
6 0.0110 0.0035
+25 V 0 1 0.0200 0.0080
6.5 0.0265 0.0096
12.5 0.0325 0.0111
18.75 0.0388 0.0127
25 0.0450 0.0143
-25 V 0 1 0.0200 0.0080
-6.5 0.0265 0.0096
-12.5 0.0325 0.0111
-18.75 0.0388 0.0127
-25 0.0450 0.0143