NI VirtualBench Calibration Procedure | © National Instruments | 59
12. Disconnect all connections and connect Channel +25 V to Volts HI and GND and Common
Floating GND to Volts LO as shown in Figure 30.
Figure 30. Cable Connections For +25 V Voltage Programming Accuracy Verification
13. Configure the DMM for the 100 V range.
14. Repeat steps 4 through 11 for the
ps/+25V channel
1 VirtualBench DC Power Supply 2DMM 3 Cabling Wire
1
3
3
HI
LO
300V
MAX
AMPS
500V MAX
HI
LO
1kV
MAX
3A, 250V
MAX
INPUT
MAX
1kV
2
HI
LO
300V
MAX
AMPS
500V MAX
HI
LO
1kV
MAX
3A, 250V
MAX
INPUT
MAX
1kV