NI VirtualBench Calibration Procedure | © National Instruments | 59
12. Disconnect all connections and connect Channel +25 V to Volts HI and GND and Common
Floating GND to Volts LO as shown in Figure 30.
Figure 30. Cable Connections For +25 V Voltage Programming Accuracy Verification
13. Configure the DMM for the 100 V range.
14. Repeat steps 4 through 11 for the
ps/+25V channel
1 VirtualBench DC Power Supply 2DMM 3 Cabling Wire