NI VirtualBench Calibration Procedure | © National Instruments | 71
10. Disconnect all connections and connect Channel +25 V to Amps HI, and Common Floating
GND to Amps LO as shown in Figure 39. Connect GND and Common Floating GND with
a jumper.
Figure 39. Cable Connections For +25 V Current Programming and Measurement
Accuracy Adjustment
11. Repeat steps 4 through 9 for the ps/+25V channel.
1 VirtualBench DC Power Supply 2DMM 3 Cabling Wire
1
3
3
2
HI
LO
300V
MAX
AMPS
500V MAX
HI
LO
1kV
MAX
3A, 250V
MAX
INPUT
MAX
1kV
HI
LO
300V
MAX
AMPS
500V MAX
HI
LO
1kV
MAX
3A, 250V
MAX
INPUT
MAX
1kV