72 | ni.com | NI VirtualBench Calibration Procedure
12. Disconnect all connections and connect Channel -25 V to Amps LO, and Common Floating 
GND to Amps HI as shown in Figure 40. Connect GND and Common Floating GND with 
a jumper.
Figure 40.  Cable Connections For -25 V Current Programming and Measurement Accuracy 
Adjustment
13. Repeat steps 4 through 9 for the ps/-25V channel.
14. Call 
PS Close Calibration with the following parameter:
• Action = 
Commit
15. Disconnect all connections.
1 VirtualBench DC Power Supply 2DMM 3 Cabling Wire
1
3