EasyManuals Logo

natus XLTEK EEG32U User Manual

natus XLTEK EEG32U
42 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #14 background imageLoading...
Page #14 background image
User and Service Manual XLTEK EEG32U Amplifier
13
Table 4 - Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications
equipment
Test
frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
Power
(W)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
385
380 390
TETRA 400
Pulse
modulation
b)
18 Hz
1,8
0,3
27
450
430 470
GMRS 460,
FRS 460
FM
c)
± 5 kHz deviation
1 kHz sine
2
0,3
28
710
704 787
LTE Band 13, 17
Pulse
modulation
b)
217 Hz
0,2
0,3
9
745
780
810
800 960
GSM 800/900,
TETRA 800, iDEN
820, CDMA
850,LTE Band 5
Pulse
modulation
b)
18 Hz
2
0,3
28
870
930
1,720
1,700 1,990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE Band
1, 3, 4, 25;
UMTS
Pulse
modulation
b)
217 Hz
2
0,3
28
1,845
1,970
2,450
2,400 2,570
Bluetooth,
WLAN, 802.11
b/g/n, RFID
2450, LTE Band
7
Pulse
modulation
b)
217 Hz
2
0,3
28
5,240
5,100 5,800
WLAN 802.11
a/n
Pulse
modulation
b)
217 Hz
0,2
0,3
9
5,500
5,785
NOTE: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME
EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a)
For some services, only the uplink frequencies are included.
b)
The carrier shall be modulated using a 50 % duty cycle square wave signal.
c)
As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual
modulation, it would be worst case.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the natus XLTEK EEG32U and is the answer not in the manual?

natus XLTEK EEG32U Specifications

General IconGeneral
Brandnatus
ModelXLTEK EEG32U
CategoryMedical Equipment
LanguageEnglish

Related product manuals