EasyManua.ls Logo

natus XLTEK EEG32U - EEG32 U Safety and Standards Conformity; Standards of Compliance and Normative References

natus XLTEK EEG32U
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
XLTEK EEG32U Amplifier User and Service Manual
8
EEG32U Safety and Standards Conformity
Standards of Compliance and Normative References
32 Channel Electroencephalograph (EEG) Headbox; model EEG32U; rated 5Vdc,0.2A
1. Type of protection against electric shock: Class II
2. Degree of protection against electric shock: Type BF
3. Degree of protection against ingress of water: IPX0
4. Degree of safety of application in the presence of a flammable anaesthetic mixture with air or
with oxygen or nitrous oxide: Equipment not suitable for use in the presence of a flammable
anaesthetic mixture with air or with oxygen or nitrous oxide.
5. Mode of operation: Continuous
6. Environmental Conditions: Normal: 10-40°C, 30-75% rH, 700-1060hPa
The EEG32U and its accessories have been designed to comply with the following national and
international standards.
Table 1 Safety Standard of Compliance and Normative References
CAN /CSA-C22.2 No. 60601-1:
08(R2013) + C2:2011
ANSI/AAMI ES60601-1:2005/(R)2012 +
C1:2009/(R)2012 and A2:2010/(R)2012
IEC 60601-1:2005 + C1:2006 and
C2:2007, Third Edition
CENELEC EN 60601-1:2006 + A1:2013
Medical electrical equipment Part 1: General requirements
for basic safety and essential performance
IEC 60601-1-6:2010, Edition 3.0
Medical electrical equipment Part 1-6: General
requirements for basic safety and essential performance
Collateral standard: Usability
IEC 62366:2007, Edition 1.0
Medical devices Application of usability engineering to
medical devices
IEC 60601-2-26:2012, Edition 3
CENELEC EN 60601-2-26L2003, Edition
2
Medical electrical equipment Part 2-26: Particular
requirements for the safety of electroencephalographs
EN ISO 80601-2-61:2011, Edition 1
Medical electrical equipment Part 2-61: Particular
requirements for basic safety and essential performance of
pulse oximeter equipment
Table 2 EMC Standard of Compliance and Normative References
IEC 60601-1-2, Edition 4.0, February
1, 2014
Medical electrical equipment Part 1-2: General
requirements for basic safety and essential
performance collateral standard: electromagnetic
compatibility requirements and tests

Related product manuals