EasyManua.ls Logo

NDB Technologies AE-150 - Page 29

NDB Technologies AE-150
61 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AE- 1 5 0 s y s t e m U s e r M a n u a l
29
Enter the transformer information (Transformer tab).
Set the communication (Communication tab).
Set the AE-150 location on the transformer tank to X=0, Y=0, Z=0 (AE-
150 position tab).
Set each sensor location from the origin point (AE-150 position tab).
Click the Controls tab.
In the Pre-set Configuration menu, select the PrPD setting of choice.
Click the AUTO-SET button. The AE-150 will set the acoustic channels
gain automatically. This process can take a few seconds.
Click on the Monitoring tab and then check the Enable checkbox.
Set the interval between two acquisitions in the Interval-Hours-Minutes
fields.
Set the Start Time and End Time.
Click the START MONITORING button to start the acquisition process.
Tip: It is recommended to let the instrument run for at least 24 hours
in monitoring mode (typically for a week). The operation requires
connecting the battery charger to the AE-150.
Monitoring results
The acquired PrPDs acquisitions will be present in the Data Analysis > Trend tab.
In the Sensors Position list, select the acquisition to be analysed.