EasyManua.ls Logo

Neuropace RNS System - Creating a Montage; Recommended Initial Montage Settings

Default Icon
112 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programming Instructions 62
RNS
®
System User Manual
Creating a Montage
1. Go to the CONFIGURE SETTINGS tab.
2. From the M
ONTAGE screen, assign an electrode to each CHANNEL (CH1 through CH4) by
selecting either a non-inverting (+) or inverting (-) input. Each electrode can be assigned to
one or two channels. Use a positive (+) input for electrodes assigned to 2 channels.
3. Assign a G
AIN for each channel by selecting a high, medium-high, medium-low, or low value.
The ECoG signal amplification may be assessed following programming by viewing the real-
time ECoGs.
Note: Using a negative (-) input for electrodes assigned to 2 channels may result in ECoG
artifacts.
Recommended Initial Montage settings
Program the initial ECoG sensing montage on the MONTAGE tab as follows:
CH 1 = Lead 1: Electrode 1 +, Electrode 2 -
CH 2 = Lead 1: Electrode 3 +, Electrode 4 -
CH 3 = Lead 2: Electrode 1 +, Electrode 2 -
CH 4 = Lead 2: Electrode 3 +, Electrode 4 -
All GAIN settings on “HIGH”
Figure 22: Montage tab

Table of Contents