Programming Instructions 67
RNS
®
System User Manual
• If ECoG storage needs to be modified, adjust the LONG EPISODE LENGTH, the MAGNET trigger
and/or the L
ENGTH OF STORED ECOG.
• The LONG EPISODE LENGTH (duration) may be adjusted (longer or shorter) to capture the
desired activity.
• The MAGNET ECoG trigger may be removed if the patient is not able to apply the magnet
before or immediately after a clinical seizure, or if the magnet-triggered ECoG is not useful.
Magnet applications will still be counted and reported in the diagnostics.
• The LENGTH OF STORED ECOG value may be increased if longer ECoGs are desired, and
decreased if a larger number of ECoGs are desired.
Figure 26: ECoG Storage tab.
Programming the Newly Selected ECoGs Storage Settings
1. Select the PROGRAM SETTINGS tab.
2. Select R
EVIEW CHANGES to view changes prior to programming.
3. Select the P
ROGRAM button.