EasyManua.ls Logo

Neuropace RNS System - Reviewing Programmer Reports

Default Icon
112 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programming Instructions 79
RNS
®
System User Manual
• Suggested initial responsive therapy settings:
• In general, stimulation should be delivered to the leads and electrodes from which the activity
of interest is observed.
• Program BURST 1 for electrodes on one lead. Program BURST 2 for electrodes on the other
lead.
• Program CURRENT to 1.0mA.
• Other parameters should be left at their default values:
• FREQUENCY = 200 Hz
• PW PER PHASE = 160 µs
• BURST DURATION = 100 ms
• Copy THERAPY 1 settings to THERAPY 2 through THERAPY 5.
• Program settings into the neurostimulator on the PROGRAM SETTINGS tab.
Figure 37: Therapy 1 tab.

Table of Contents