Programming Instructions 79
RNS
®
System User Manual
• Suggested initial responsive therapy settings:
• In general, stimulation should be delivered to the leads and electrodes from which the activity
of interest is observed.
• Program BURST 1 for electrodes on one lead. Program BURST 2 for electrodes on the other
lead.
• Program CURRENT to 1.0mA.
• Other parameters should be left at their default values:
• FREQUENCY = 200 Hz
• PW PER PHASE = 160 µs
• BURST DURATION = 100 ms
• Copy THERAPY 1 settings to THERAPY 2 through THERAPY 5.
• Program settings into the neurostimulator on the PROGRAM SETTINGS tab.
Figure 37: Therapy 1 tab.