EasyManua.ls Logo

New Focus 3501 - 4 Quick Start

New Focus 3501
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8
Before operating the Model 3501 Optical Chopper, please review the previous
section to make sure the chopper is ready for use.
1. Mount the chopper head securely on an optical bench.
2. Mount the 42/30 wheel on the head. Avoid bending the wheel. The wheel
should be free to turn without contact to the optical sensor or nearby
equipment.
3. Connect the cable between the chopper head and the connector marked
MOTOR on the back panel of the control unit.
4. Check that the correct line voltage has been selected and the proper fuse
inserted into the mains power entry module at the back panel of the con-
trol unit. Connect the instrument to the power mains.
5. After insuring the safety of nearby personnel, switch on the instrument.
6. Recall instrument set-up “0” (factory default setting). Press SET until
“RECALL” is lit. Press the up/down arrow keys until “0” appears in the
display window. Press the left or right arrow key to recall set-up “0”.
7. The instrument will lock at 84 Hz within 1 minute.
8. To choose a particular chopping frequency, press the Set key until the
FREQ LED is lit, and then use the up/down arrows keys to adjust the
chopping frequency.
IV
Quick Start
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com