26
FR*?
Frequency query.
The FR* command is used to query the following frequencies associated with
the Measure key on the chopper front panel: F
sync
, F
outer
, OUT1, and OUT2.
Data is returned in Hz.
FR1? - F
sync
FR2? - F
outer
FR3? - OUT1
FR4? - OUT2
The chopping frequency is 1.23 kHz, and the chopper is in NORMAL mode.
FR1?
—>FR11230.0
FR2?
—>FR21230.0
FR3?
—>FR36150.0 (The frequency at OUT1 is 6.15 kHz.)
Description
Command
Explanation
Example