VII. Technical specifications
65
VII. Technical specifications
1. When the microscope (TE2000-E, TE2000-U, TE2000-S) is used in combination
with T-DH dia-illuminator 100W (for 12V100W lamp)
Combination Microscope + T-DH dia-illuminator 100W + TE2-PS100W power supply
Dimensions
568(W) × 727(D) × 703(H) mm
Weight 41 Kg
Opto-mechanical specification
Objectives: CFI60
Eyepieces: Field Number 22
Nosepiece : 6-place
Focusing:
Stroke 10 mm
Coarse 4.9 mm/rev.
Fine 0.1 mm/rev.
Fine scale 0.05 μm (TE2000-E), 1 μm (TE2000-U,S)
Input ratings of T-DH dia-illuminator
100W
DC 12 V, 100 W
Lamp ratings 12V100W halogen lamp
Lamp type Halogen lamp (OSRAM HLX 64623 or PHILIPS 7724I)
Average lamp lifetime 2,000 hrs.
TE2-PS100W power supply
Input ratings: AC 100-240 V, 50/60 Hz, 2.4 A
Voltage fluctuation: ±10%
Output ratings: DC 12 V, 100 W, 8.4 A
Built-in fuse ratings: 250 V, T4A
Protection class: Class 1
Others: UL Recognized product, GS approved
Operating environmental conditions
Altitude: 2,000 m max.
Temperature: 0 to 40ºC
Relative humidity: 85% max. (no condensation)
Pollution degree: Degree 2
Indoor use only
Storage and transport environmental
conditions
Temperature: -20 to +60
Humidity: max. 90% RH (no condensation)
Installation category Category II