Control Test
e Control Test checks that the System is working
correctly and testing technique is good. Use ONLY
TRUEyou Control Solution to perform Control
Tests. Perform Control Tests:
For practice before using the System for the rst
time,
When opening a new vial of Test Strips,
Occasionally as a vial of Test Strips is used,
If a Test Strip vial has been le opened or le in
extreme heat, cold, or humidity,
Whenever a check on performance of the System is
needed,
If results seem unusually high or low,
If Meter damage is suspected (Meter was dropped,
crushed, wet, etc.).
Performing Control Tests with more than one level
of Control Solution is recommended to ensure that
your System is working properly. ree levels of
TRUEyou Control Solution are available. Contact
place of purchase or use the number on the cover of
this Booklet for more information on how to obtain
dierent levels of Control Solution.
Ranges printed on Test Strip vial label are
for Control Test results only and are not
suggested levels for your blood glucose. Do
not drink Control Solution.