Control Solution Care
Write date opened on Control Solution label.
Discard if either 3 months aer rst opening or
aer date printed next to on bottle label has
passed, whichever comes rst.
Store at 2°C-30°C (room temperature).
DO NOT REFRIGERATE
OR FREEZE.
Aer use, wipe bottle tip clean and recap tightly.
Test Strip Care
Store Test Strips in original vial only. Do not
transfer old Test Strips to new vial or store Test
Strips outside of vial.
Write date opened on Test Strip vial. Discard
unused Test Strips from vial if either 4 months aer
rst opening or aer date printed next to on vial
label has passed, whichever comes rst. Use of Test
Strips past either date may give incorrect results.
Close vial immediately aer removing Test Strip.
Store in a dry place below
30°C (room temperature).
DO NOT REFRIGERATE
OR FREEZE.
Do not reuse Test Strip. NEVER wipe Test Strips
with water, alcohol or any cleaner. DO NOT
attempt to clean and re-use Test Strips. Reuse of
Test Strips will cause inaccurate results.
Do not bend, cut or alter Test Strips in any way.
2°C
30°C
1°C
30°C