EasyManua.ls Logo

NIPRO Diagnostics TRUEyou - How to Test Control Solution

NIPRO Diagnostics TRUEyou
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
17
How to Test Control Solution
Use ONLY TRUEyou Control Solution
with the TRUEyou Meter.
1. Check supplies. See Getting Started.
2. Allow Control Solution, vial of Test
Strips and Meter to adjust to room
temperature (20°C-25°C).
Note: Running a Control Test at
temperatures outside the range
listed above may cause Control
Solution to read as a blood test.
3. Wash hands. Dry thoroughly.
4. Gently swirl or invert Control Solution bottle to
mix. DO NOT SHAKE!
5. Remove one Test Strip from vial. Close Test
Strip vial immediately. Use Test Strip quickly
aer removal from vial.
Note: If Test Strip has been out of the
vial too long before testing, an
error message appears upon
insertion of the Test Strip into the
Meter. Release and discard old
Test Strip. Use new Test Strip for
testing.
6. Insert Test Strip into Meter. Meter turns on
and shows blinking Drop Symbol and time. Do
not remove Test Strip.
Contacts
Facing Up
May 30, 2014
Tyou_OB_Guts_E4GBY03FDr20.indd 17 11/8/13 1:43 PM

Related product manuals