Display What it Means What to Do
Software Error Inside the USA, call 1-800-
681-7390. Outside the
USA, contact your Nova
supplier.
System Hardware Error Inside the USA, call 1-800-
681-7390. Outside the
USA, contact your Nova
supplier.
Operating Temperature
Error
Monitor is outside the
requiredtestingtem-
peraturerangeof57°
to104°F(14°to40°C).
Move the monitor to a
warmer or cooler area
and wait a few minutes.
Used Strip Error: Used or
damaged strip.
Retest with a new strip.
Blood Sample Error
Incorrect application of
blood sample or control
solution onto the test strip,
or the test strip may be
damaged. Review your
samplingtechnique.
Displays, Meanings, Actions
Nova Max Plus Monitor OG PN46171D1 EN.indd 45 9/5/12 3:22 PM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nova Max PLUS and is the answer not in the manual?

Nova Max PLUS Specifications

General IconGeneral
Test Time5 seconds
HCT Range30 - 55%
Operating Temperature50-104°F (10-40°C)
Operating Humidity10-90% RH
Blood Sample Size0.5 µL