46
Displays, Meanings, Actions
Display What it Means What to Do
Monitor
does not
turn on after
inserting a
test strip.
• Teststripisinserted
upside down or not
completely in.
• Batteryisdead.
• Batteryisinstalledin-
correctly or there is no
battery in the monitor.
Insert the test strip
correctly with the Nova
name and white tip fac-
ing up and out.
Replace the battery.
Check that the battery is
correctly installed with
the “+” sign facing you.
Inside the USA, call 1-800-
681-7390. Outside the
USA, contact your Nova
supplier.
Monitor
does not
begin test
countdown
after apply-
ing a blood
sample.
• Notenoughblood
sample.
• Sampleappliedafter
monitor automatically
turned o.
• Teststripmaybe
damaged.
• Monitormaynotbe
working properly.
Repeat the test with a
new test strip.
Repeat the test with a
new test strip.
Repeat the test with a
new test strip.
After 3 attempts, call
1-800-681-7390 (inside
the USA). Outside the
USA, contact your Nova
supplier.
Nova Max Plus Monitor OG PN46171D1 EN.indd 46 9/5/12 3:22 PM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nova Max PLUS and is the answer not in the manual?

Nova Max PLUS Specifications

General IconGeneral
Test Time5 seconds
HCT Range30 - 55%
Operating Temperature50-104°F (10-40°C)
Operating Humidity10-90% RH
Blood Sample Size0.5 µL
Measuring Range20-600 mg/dL