41Emitted Radiation & Electromagnetic Compatibility
30 Emitted Radiation & Electromagnetic Compatibility
The Optune
®
Treatment Kit and the accompanying battery charger (ICH9100) and power supply (SPS9100) need special
precautions regarding EMC and need to be installed and put into service according to the EMC information provided
below.
Portable and mobile RF communications equipment can aect the Optune Treatment Kit System and the
accompanying battery charger.
The Optune device should be used with the following cables and additional parts only:
1. CAD9100 connection cable
2. INE9000 INE transducer array (Sterile)
3. IBH9100 battery
4. SPS9100 power supply
5. ICH9100 charger
6. Unshielded AC mains cables for indoor use only with a maximal length of less than 3.0 m
The use of accessories, parts and cables other than those specified, may result in increased EMISSIONS or decreased
IMMUNITY of the Optune Treatment Kit.
Table 1 – Guidance and MANUFACTURER’S declaration – ELECTROMAGNETIC EMISSIONS – for
all ME EQUIPMENT and ME SYSTEMS
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
Optune Treatment Kit is intended for use in the electromagnetic environment specified below. The customer
or the user of Optune Treatment Kit should ensure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11 Group 1 Optune Treatment Kit uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11 Class B Optune Treatment Kit is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions IEC 61000-3-2 Class A
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Complies
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The ICH9100 charger and the SPS9100 power supply are intended for use in the electromagnetic environment
specified below. The customer or the user of the ICH9100 charger and the SPS9100 power supply should ensure
that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11 Group 1 The ICH9100 charger and the SPS9100 power supply
use RF energy only for their internal function. Therefore,
their RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class B The ICH9100 charger and the SPS9100 power supply
are suitable for use in all establishments, including
domestic establishments and those directly connected
to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2 Class A
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Complies