43
30 Emitted Radiation & Electromagnetic Compatibility
Emitted Radiation & Electromagnetic Compatibility
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The ICH9100 charger and the SPS9100 power supply are intended for use in the electromagnetic environment
specified below. The customer or the user of the ICH9100 charger and the SPS9100 power supply should ensure
that it is used in such an environment.
Emissions test IEC 60601
Test level
Compliance level Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV
N/A
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line to line
±2 kV line to earth
±1 kV line to line
±2 kV line to earth
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input
Power frequency
(50/60 Hz)
magnetic field
<5% UT
(>95% dip in UT) for 0,5 cycle
40% UT
3 A/m
<5% UT
(>95% dip in UT) for
0,5 cycle
40% UT
3 A/m
Mains power quality should be that
of a typical commercial or hospital
environment.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
NOTE UT is the AC mains voltage prior to application of the test level = 120V and 230V
Normal operation: The Optune
®
Treatment Kit is working properly when the blue LED surrounding the TTFields button
are lit and no notification signal sounds. The ICH9100 charger is working properly when all the LEDs are lit. The SPS9100
power supply is working properly when the blue LEDs surrounding the TTFields button on Optune Treatment Kit are lit
and no notification signal sounds.