(3) Calibration
This function is to decrease fluctuation in the rotation speed of the motor handpiece and the
difference in torque by the contra angle head.
1. Lubricate the contra angle head with NSK PANA SPRAY/PANA SPRAY Plus (option). (Refer
to the “7 (1) Lubricating Contra Angle Head.”)
2. Turn on the power.
3. Hold down UP/DOWN key simultaneously for more than one second.
4. The LCD panel displays “ ” with the alarm sound.
5. Attach the contra angle head to the motor handpiece and press the ON/OFF key. (Remove
the bur or TEST bur.)
6. If the motor handpiece begins to rotate, leave it as it is until it stops.
7. This process ends, if the rotation stops and the LCD panel display returns to its original
state.
8. If you want to stop this process, turn off the power.
(4) Alarm Sound Volume Control
1. Setting alarm sound
You can turn ON/OFF the alarm (load alarm) which sounds near the torque limit value when the
motor handpiece rotates and the alarm during auto reverse (auto reverse sound).
1. Turn off the power
2. While pressing SELECT key, keep pressing POWER key one second or more.
3. The alarm mark and ON or OFF on the LCD panel, and the sound volume changes.
4. Press the SELECT key to select ON or OFF for alarm sound.
5. Display on the liquid crystal panel (LCD) returns to the original state if no operation is
performed for a while.
(1) Program
You can change any preset values and have them memorized to your desired settings. (rotation
speed, torque limit value, gear ratio and auto reverse mode)
1. Press the PROGRAM key until it turns to the program number which you want to have
memorized.
2. Adjust the rotation speed, torque limit value, gear ratio and auto reverse mode by UP/DOWN
key according to your needs.
3. Hold down the PROGRAM key for more than one second. When the alarm sounds the
settings have been memorized.
(2) Initialization of Program (Factory-set configuration)
The program can be returned to the original state configured at the time of factory shipment, if
setup becomes confusing.
1. Turn off the power, when the power is on.
2. Hold down the PROGRAM key and ON/OFF key for more than one second, while pressing
the POWER key at the same time.
3. Release the keys when the LCD panel indicates “ ”, and re-press the PROGRAM key
while this is displayed.
4. Initialization is completed, when “ ”is displayed.
6. Convenient Function
(6) Completion of Medical Treatment
When the treatment is completed, return the motor handpiece to the handpiece stand and hold
down the POWER key for more than one second to turn off the power.
Auto Power Off
If the key is not operated or approximately 10 minutes passes during rotation with no-load, the
power is automatically turned OFF to save energy and prevent improper operation. However, if
load is applied during rotation of the motor handpiece, the power is not turned OFF even during
no operation.
LAST MEMORY
The last setting used before the power is turned OFF is memorized. The setting when the
power is turned ON next is as that immediately used before the power has been turned OFF.
• The program cannot be memorized while the motor handpiece is in motion.
• The program is not memorized unless the PROGRAM key is held down for more than
one second. If the program number is changed by the PROGRAM key, the initially
memorized preset values remain. (Cancel function)
NOTICE
If this function is used, all programs will disappear and return to the originally set values.
Record the present program details if required before you carry out this operation.
NOTICE
NOTICE
• This function does not work unless remaining battery capacity is sufficient.
• Perform calibration after cleaning the contra angle head. Residual contamination on
the rotating shaft will impair correct measurement.
• Do not touch or apply a load to the rotating shaft of the contra angle head. It will
obstruct accurate measurement.
• This function cannot completely absorb the individual difference of the actual motor
handpiece and contra angle head.
NOTICE
The volume setting is memorized automatically to the last setting and remains
unchanged even if the power is turned off.