Sterilize the contra angle head only.
For the sterilization method, we recommend the autoclaving. Sterilization is required first time
you use and after each patient as noted below.
Autoclaving
1. Brush the dirt off the surface of the contra angle head, and wipe it off with a cotton cloth
moistened with surgical spirit. Do not use a metal brush.
2. Lubricate the head with the PANA-SPRAY/PANA-SPRAY Plus. (Refer to the “7 (1) Lubricating
Contra Angle Head.”)
3. Insert the head into an autoclave pouch and seal it.
4. Autoclavable up to a max. 135°C (275°F).
ex.) Autoclave for 20 min. at 121°C (250°F), or 15 min. at 132°C (270°F).
5. Keep the handpiece in the autoclave pouch to keep it clean until you use it.
* Sterilization at 121°C for more than 15 minutes is recommended by EN13060 or
EN ISO17665-1.
8. Sterilization
(1) Lubricating Contra Angle Head
• Lubricate the contra angle head only.
• Apply PANA SPRAY/PANA SPRAY Plus after each use and/or before each calibration and
autoclaving
1. Lubricate it before each autoclaving or once a
day if you do not autoclave, using
PANA-SPRAY/PANA-SPRAY Plus lubricant.
Screw the F type spray nozzle onto the
Pana-Spray nozzle by approx. 10 turns.
2. Insert the F type spray nozzle into the rear
part of the contra angle head and lubricate the
head for 2-3 seconds. If the contra angle head
does not push into nozzle enough, oil may not go around into the handpiece and it may flows
backward.
(2) Cleaning Motor Handpiece
When the motor handpiece becomes dirty, wipe it off with a cotton cloth moistened with
rubbing alcohol.
(3) Cleaning Charger
When the charger becomes dirty, wipe it off with a cotton cloth moistened with surgical
solution.
7. Cleaning
• Do not heat sterilize the motor handpiece.
• Do not wipe with, or clean or immerse in, high acid water or sterilizing solutions.
CAUTION
• Do not lubricate the motor handpiece.
• Before mounting the lubricated contra angle head to the motor handpiece, wipe off
excess oil. Stand it on its end or lean it in the proper position for gravity draining.
Mount it after excess oil has been completely drained.
• Hold the contra angle head securely to prevent it from flying off by the pressure of the
spray.
• Hold the spray can upright.
• When cleaning the motor handpiece, do not use any solvent such as benzene and,
thinner.
CAUTION
Fig. 9
When the contra angle head is removed from the
motor handpiece for battery charging,
lubrication, or sterilization, mount the motor cap
onto the motor handpiece to prevent debris from
entering.
9. Motor Cap
Fig. 10