English
Battery charger
Power cord
Contra Angle Handpiece
(Included in the MP-F16R SET)
F-type spray nozzle
Motor cap
Motor Handpiece
8 9
2. Specification
• Ergonomic design and compact body.
• “6 position type head” allows direction of contra angle head to be changed and the ON/OFF
key to be set to a position where it can be easily activated.
• The larger ON/OFF key enhances operability (compared with conventional products).
• Operates continuously for approx. 1.2 hour at rated load. (Dependent upon use conditions.)
• The liquid crystal panel enhances user friendliness.
• Memory for up to 5 programs .
• The auto reverse function is activated depending on the load. A wide variety of functions such
as “AUTO REVERSE”, “AUTO STOP” and “AUTO REVERSE OFF” are available.
• The motor handpiece can be turned on and off by pressing the ON/OFF key. Alternatively the
unit can be operated by holding down the ON/OFF key during operation and letting go of the
ON/OFF key to stop the handpiece.
• The motor handpiece softly starts. Since the rotation briefly stops before changing direction
there are no vibrations and shocks during the change of directions.
• Energy saving. The power of the motor handpiece automatically turns off when there are no
operations for 10 minutes. (Auto power-off function)
• The feedback circuit, which keeps rotation at a constant speed even when the load on the
motor handpiece changes, is built in.
• Non-contact charger prevents improper charging due to deterioration of metallic terminal.
•
In spite of non-contact charging, this has realized rapid charging, shortening the time for charging.
• The contra angle heads provided for this product are all autoclavable at 135ºC (excluding the
motor handpiece).
1. Features
3. Name of each part
Symbols
This conforms to CE European Directive of “Medical equipment directive 93/42/EEC.”
TUV Rhineland of North America is a Nationally Recognized Testing Laboratory (NRTL) in
the United States and is accredited by the Standards Council of Canada to certify
electro-medical products with Canadian National Standards.
Follow the waste of electric and electronic equipment (WEEE) Directive (2002/96/CE) to
dispose of the product and accessories.
Class II equipment
• The motor handpiece is designed for commercially available AAA nickel metal hydride
batteries (rechargeable).
• The motor handpiece consumes electricity very slightly even when the power is off. In
addition, fully-charged rechargeable batteries, in general, discharge gradually over time
even though it is not used. It is recommended to recharge the batteries just before use.
• When the motor handpiece automatically stops by detecting a low battery voltage, leaving
it for a while and turning on the power again may not detect the low voltage immediately.
This is not a failure, but due to battery characteristics. Since the voltage drop does not
coincide with the remaining battery capacity, consider it only as a yardstick.
•
Recharge rechargeable batteries after they deplete as much as possible. Repeating short
time use and subsequent recharging may shorten their operating time due to a “memory
effect.” Batteries may recover from the memory effect after fully discharging and then fully
charging and repeating this procedure a few times.
• Eventually completely discharged batteries cannot be charged and need to be replace with
new ones.
• User shall be responsible for operation, maintenance and inspection.
• No special training is required for this device.
NOTICE
Refer to the Operation Manual.
Type B conforming component
Manufacturer
Battery Charger
Model
Input Voltage
Input Power
Charging Time
NE233
AC120/240V±10%
50/60Hz
15VA
Approx. 90 min.
Motor Handpiece
Model
Input Voltage
Input Power
EM10M2
DC2.4V±20%
0.3VA
Use environment
Temperature
Humidity
Atmospheric pressure
10 - 40 ˚C
10 - 75 %
(Non condensing)
700-1060 hPa
Store environment
Temperature
Humidity
Atmospheric pressure
-10 - 50 ˚C
10 - 80 %
(Non condensing)
500-1060 hPa