Guidance and manufacturer's declaration - electromagnetic immunity 
The ENDO-MATE TC2 is intended for use in the electromagnetic environment specified below. The 
customer or the user of the ENDO-MATE TC2 should assure that it is used in such an environment. 
NOTE: Ut is the a.c. mains voltage prior to application of the test level. 
Immunity test  IEC60601 test level 
Electromagnetic environment - guidance 
Electrostatic 
discharge (ESD) 
IEC61000-4-2 
 
 
Electrical fast 
transient/burst 
IEC61000-4-4 
 
 
Surge 
IEC61000-4-5 
 
 
Voltage dips, 
short 
interruptions and 
voltage variations 
on power supply 
input lines 
IEC61000-4-11 
 
 
 
 
 
 
Power frequency 
(50/60Hz) 
magnetic field 
IEC61000-4-8 
±6kV contact 
±8kV air 
 
 
 
±2kV for power 
supply lines 
±1kV for 
input/output 
 
±1kV line(s) to line(s) 
±2kV line(s) to earth 
 
 
<5% Ut 
(>95% dip in Ut) 
for 0.5 cycle 
40% Ut 
(60% dip in Ut) 
for 5 cycles 
70% Ut 
(30% dip in Ut) 
for 25 cycles 
<5% Ut 
(>95% dip in Ut) 
for 5 secs 
 
3 A/m 
 
Compliance level 
±6kV contact 
±8kV air 
 
 
 
±2kV for power 
supply lines 
 
 
 
±1kV line to line 
±2kV lines to earth 
 
 
<5% Ut 
(>95% dip in Ut) 
for 0.5 cycle 
40% Ut 
(60% dip in Ut) 
for 5 cycles 
70% Ut 
(30% dip in Ut) 
for 25 cycles 
<5% Ut 
(>95% dip in Ut) 
for 5 sec 
 
3 A/m 
 
Floors should be wood, concrete 
or ceramic tile. If floors are covered 
with synthetic material, the relative 
humidity should be at least 30%. 
 
Mains power quality should be that 
of a typical commercial or hospital 
en
vironment. 
 
 
Mains power quality should be that 
of a typical commercial or hospital 
environment. 
 
Mains power quality should be that 
of a typical commercial or hospital 
environment. If the user of the 
ENDO-MATE TC2  requires 
continued operation during power 
mains interruptions, it is 
recommended that the ENDO-MATE 
TC2  be powered from an 
uninterruptible power supply or a 
battery. 
 
 
 
Power frequency magnetic fields 
should be at levels characteristic of a 
typical location in a typical 
commercial or hospital environment. 
Guidance and manufacturer's declaration - electromagnetic immunity 
The ENDO-MATE TC2 is intended for use in the electromagnetic environment specified below. The 
customer or the user of the ENDO-MATE TC2 should assure that it is used in such an environment. 
NOTE 1  At 80MHz and 800MHz, the higher frequency range applies. 
NOTE 2  These guidelines may not apply in all situations. Electromagnetic propagation is affected 
by absorption and reflection from structures, objects and people. 
a 
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones 
and land mobiles radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be 
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF 
transmitters, an electromagnetic site survey should be considered. If the measured field strength in 
the location in which the ENDO-MATE TC2 is used exceeds the applicable RF compliance level 
above, the ENDO-MATE TC2 should be observed to verity normal operation. If abnormal 
performance is observed, additional measures may be necessary, such as reorienting or relocating 
the ENDO-MATE TC2. 
b  Over the frequency range 150kHz to 80MHz, field strengths should be less than 3 V/m. 
Immunity test 
IEC60601 test level  
Electromagnetic environment - guidance 
 
 
 
 
 
 
 
 
 
 
 
Conducted RF 
IEC61000-4-6 
 
Radiated RF 
IEC61000-4-3 
 
 
 
 
 
 
 
 
 
 
 
3Vrms 
150 kHz to 80MHz 
 
3V/m 
80MHz to 2.5 GHz 
 
 
 
Compliance level  
 
 
 
 
 
 
 
 
 
 
 
3Vrms 
 
 
3V/m 
 
Portable and mobile FRO 
communications equipment should be 
used no closer to any part of the 
ENDO-MATE TC2, including cables, 
than the recommended separation 
distance calculated from the equation 
applicable to the frequency of the 
transmitter. 
 
Recommended separation distance 
 
d = 1.2  P 
 
d = 1.2  P  80MHz to 800MHz 
d = 2.3  P  800MHz to 2.5GHz 
 
Where P is the maximum output power 
rating of the transmitter in watts (W) 
according to the transmitter 
manufacturer and d is the 
recommended separation distance in 
meters (m). 
Field strengths from fixed RF 
transmitters as determined by an 
electromagnetic site survey, should be 
less than the compliance level in each 
frequency range. 
Interference may occur in the vicinity of 
equipment marked with the following 
symbol: