8-3 Disinfection
Manual Disinfection (External Surface)
Wipe the surface with cloth soaked in disinfectant solution
such as alcohol.
Automatic Disinfection (External and Internal Surfaces)
(Only for FX15m / FX23m / FX25m / FX75m)
Use a thermo-disinfector in accordance with ISO 15883
(EN ISO 15883).
Refer to the thermo-disinfector manual.
8-4 Lubrication
NSK PANA SPRAY Plus / PANA SPRAY
Apply NSK PANA SPRAY Plus / PANA SPRAY every time after
each use and/or before autoclaving.
1) Attach the Spray Nozzle into the Spray Port nozzle on the
can.
2) Insert the Spray Nozzle into the rear of the hand
piece. Hold
the handpiece and spray for approximately 2-3 seconds.
Apply lubricant until it expels from the handpiece head for
at least 2 seconds (Fig. 8).