iCare Quick Operation Guide
Be sure to prime iCare with the Maintenance Oil when using the product for the first time.
Attaching Handpiece
w182x h257 表
Be sure to prime iCare with the Maintenance Oil when using the product for the first time or after draining the oil from
the oil tank in accordance with Operation Manual "6. Oil priming". Lack of oil priming may reduce the performance of the product.
For Contra Angle and Air Turbine Handpieces point the Bur
Insertion Hole towards the door and close the door (Fig. 4).
Mist Filters absorb the oil coming out of the handpiece.
When connecting Air Turbine
① Remove the bur from the handpiece.
② Open the iCare Door, insert the adaptor (option) to the shape of the air turbine joint,
and tighten the nut securely.
③ Insert the air turbine into the adaptor (Fig. 1).
④ Push and pull the handpiece to be sure that it is securely attached.
⑤ Point the bur insertion hole toward the door and close the door securely.
*When connecting a Midwest 4 Hole air turbine, insert the air turbine to the shape
of the air turbine joint, and tighten the nut securely.
*When connecting Straight Handpiece, insert the supplied Test Bur into the handpiece
and lock the chuck by turning the Bur Lock Ring (Fig.2).
① Remove the bur from the handpiece.
② Open the iCare door and attach the handpiece to the E-Type handpiece joint until
it locks (Fig. 3).
③ Push and pull the handpiece to be sure that it is securely attached.
④ Point the bur insertion hole toward the door and close the door securely.
Air Turbine
Fig.1 Fig.2 Fig.3
Air Turbine Joint
Tighten
Nut
Adaptor
When connecting Contra Angle Handpiece
E Type handpiece
Bur Lock Ring Test bur
E-Type
handpiece Joint
Fig.4
•
Do not fill Maintenance Oil in a room where there is a risk of explosion or fire.
•
Do not overfill Maintenance Oil in order to avoid accidental oil leaks. Pour slowly.
Cleaning the iCare
NOTICE
Filling with Maintenance Oil
2014.03.20
OM-E0693E
① Open the Cover.
② Remove the Oil Fill Cap by turning it counter clockwise.
③ Remove the cap for Maintenance Oil and install the Oil Filling Nozzle.
④ Fill with Maintenance Oil slowly (Fig. 7).
⑤ When filling is complete, securely attach the Oil Fill Cap by turning it clockwise.
•See Operation manual
"
9-3〜9-8
"
about the replacing of Mist Filter, Oil Absorber Sheet, O-ring, or changing Fuses.
•Remove oil in the oil tank following Operation Manual
"
9-6Oil removal from the oil tank
"
, before relocating or requesting product service.
•Never use solvent such as benzine or thinner for cleaning.
•Do not leave an alcohol-immersed cloth in contact with the Unit for a long period.
Bleaching, deformation or failure of the Unit may occur.
•The door can easily be detached when necessary and be cleaned with tap water
(See Operation Manual "9-2 Detaching and reattaching the Door").
Drain the Air Compressor and Air Filter at least once a week as required. Moisture from
the tank could mix with the Maintenance Oil.
Maintenance of iCare
① Turn OFF (O side) the Power Switch.
② Remove the AC power cord from Unit.
③ Wipe the surface using a cloth moistened with water, warm water or soapy water
(neutral detergent), then again wipe using another cloth moistened with disinfectant
alcohol.
Draining the Air Filter
① When water accumulates in the Air Filter case,
press the Drain Button at the bottom of the case
to drain the Air Filter (Fig. 8).
② In addition to the Air Filter, drain the air supply
attached to this device (Air Compressor) at least
once a week.
Fig.7
Fig.8
Drain Button