9
English
_ Cleaning and Disinfection
Perform cleaning and thermal disinfection using the washer-disinfector under the following conditions:
Temperature Time Water quality Detergent
Pre-cleaning < 25°C 3 min Tap water -
Cleaning 40 - 60°C ≥ 5 min
Distilled water
Deionized water
Demineralized water
Neutral enzymatic detergent for medical
purposes (pH 7.0 - 8.0), or alkaline detergent
for medical purposes (pH ≤12)
Rinsing > 10°C ≥ 1 min -
Thermal disinfection 90 - 93°C 5 min -
Drying Set the washer-disinfector according to the instruction of the manufacturer.
* Use neutral enzymatic detergent for medical purposes for P200-WPD.
* Detergent used for validation:
Neutral enzymatic detergent: neodisher MediZym (DR. WEIGERT)
Alkaline detergent: neodisher SeptoClean (DR. WEIGERT)
_ Drying
1
Remove moisture on the surface of the product using a dry
cloth, air gun, etc.
2
Visually check that all contamination is removed.
In case there is residual contamination, repeat the procedures
of Pre-cleaning through Drying until all contamination is
removed.
Proceed to Checking (see “6 2 Checking”).
6 – 2 Checking
1
Check for any abnormalities or damage on the outer surface
and for any damage or deformation on the Pins in the Plug.
2
Check that the Motor can be securely connected to the Control
Unit.
3
Depress the Pedal of the Foot Control to check for any abnormal
vibration, noise or heating in the Motor.
Check on the Main Panel of the Control Unit that the rotational
speed can reach the maximum speed you have set.
4
Check that Forward, Reverse, and Oscillate (P200-WPD only)
Modes operate normally.
5
Mount the Attachment to the Motor to check its secure
connection.
Proceed to Packaging / Sterilization (see “10 Packaging /
Sterilization”).
CAUTION
Do not lubricate the Motor. It may cause failure.

Table of Contents

Other manuals for NSK Primado2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the NSK Primado2 and is the answer not in the manual?

NSK Primado2 Specifications

General IconGeneral
BrandNSK
ModelPrimado2
CategoryMedical Equipment
LanguageEnglish

Related product manuals