22
Cleaning / Checking the Hood, Beak, Duraguard, Rotatable Duraguard, Wrench and Adaptor
_ Drying
1
Remove moisture on the surface of the product using a dry
cloth, air gun, etc.
2
Visually check that all contamination is removed.
In case there is residual contamination, repeat the procedures
of Pre-cleaning through Drying until all contamination is
removed.
Proceed to Checking (see “8 2 Checking”).
8 – 2 Checking
Check the Hood, Beak, Duraguard or Rotatable Duraguard with the same procedure described in "Check before Operation".
Check for any scratches, damage or deformation of the Wrench and Adaptor.
Proceed to Packaging / Sterilization (see “10 Packaging / Sterilization”).
* Lubrication is unnecessary for Hood, Beak, Duraguard, Rotatable Duraguard, Wrench and Adaptor.
NOTICE
To perform "Check before Operation", refer to OM-SH0913MA Primado2 OPERATION MANUAL Attachments.