Cleaning / Lubrication / Checking of the Attachment, Slim Attachment Hub and Slim Tube
English
7 – 1 – 1 Manual Cleaning
Manual Cleaning using the Cleaning Tools
Use the Cleaning Tools according to the following system figure.
*1: With O-ring (Order Code: D0312074080)
*2: With O-ring (Order Code: D0312150150)
Attachment
Cleaning Adaptor SM
Slim Attachment Hub
Cleaning Adaptor 1TSlim Tube
Cleaning Adaptor HM
*1
Attachment
Cleaning Adaptor WPD
*2
Attachment
Cleaning Drying
Luer-lock syringe
Cleaning Adaptor G Air Gun
_ Pre-cleaning
Remove contaminants on the surface with a cloth or soft bristle brush under running tap water.
For products with moving parts, perform cleaning while moving all parts (see “2 Moving Parts of Each Product”).
≤ 38°C, ≥ 3.5 L/min, ≥ 30 sec.
Water quality: Drinking water
For Slim Attachment Hub, perform this step with the Slim Tube removed.
_ Cleaning
1
Turn the Twist Collet to the closed position (if applicable).
2
Mount the Cleaning Adaptor.
Pull the Cleaning Adaptor to make sure that it is securely
connected.
3
Screw the disposable luer-lock syringe (hereafter referred to as
syringe) into each Cleaning Adaptor of the Attachment.
4
Prepare cleaning solution in the cleaning tank under the
following conditions (prepare the cleaning solution according to
the instructions of the detergent manufacturer).
Temperature Water quality Detergent
45°C
Distilled water
Deionized water
Demineralized
water
Neutral enzymatic
detergent for medical
purposes (pH 7.0 - 8.0)
* Detergent used for validation:
neodisher MediZym (DR. WEIGERT)