5-8 Mounting the Tip
1) First, lightly screw in the Tip with by hand. (Fig. 15)
2) Match the square shape of the Tip with the hole of the Tip Wrench, then insert the Tip into the hole. (Fig. 16)
3) Rotate the wrench in the (clockwise) direction, (Fig. 17 and 18) until it makes a click sound and cannot rotate any
further.
* To remove the Tip, rotate the wrench in the (counter-clockwise) direction shown in Fig. 17 and Fig. 18.
Fig. 16Fig. 15
Fig. 18Fig. 17
TightenLoosen
Tighten
Loosen
CAUTION
ŘWhen attaching or removing a Tip, which is longer than the Tip Wrench, be careful that the protruding Tip
does not cause an injury.
ŘDo not use a Tip which has been damaged, bent or corroded as it will be weakened and may break while
in use.
ŘTo avoid bending of the Tip thread or disengaging of the Tip, always make certain that the Tip is inserted
at the correct angle and firmly tightened.
ŘIf debris is found on the Tip thread clean the Tip immediately so as to avoid possible weak vibration
ŘDo not attempt to sharpen or bend the Tip as this may cause the Tip to break during operation, or the
vibration may be weakened.
ŘWhen Tip become worn vibration may weaken or the Tip may suddenly break. If a weak Tip is detected
replace with a new one.
ŘTips need to correctly inserted and tightened using the Tip Wrench. If the Tip is not tightened sufficiently
vibration may be weak.
ŘWhen inserting a Tip always wear surgical gloves and ensure the Tip, Handpiece and Tip Wrench are in an
hygienic condition.
ŘBefore connecting or disconnecting the Handpiece Cord or Irrigation Tube make sure to remove the Tip
first to avoid injury by the Tip.
ŘTip Wrench will wear with frequent use time and may need annual replacement.
ŘUse only genuine NSK Tips. Other brand Tips may cause malfunction, damage or injury:
- Poor vibration when an incompatible thread has been forcibly fitted.
- The patient may swallow pieces of broken non genuine Tips.
- Damage or premature wear of the Handpiece thread.