- Page 17 -
Nu-Tek Levator Elite User Manual
EMG Parameters
Note:
1. In Therapy Mode, you can change all the EMG Biofeedback settings.
2. Electrodes placed on the lower abdominals, legs, arms and back, face,
buttock or pelvic area use WIDE lter band setting.
All other areas i.e.upper abdominals, chest, shoulder, upper arms and back
use NARROW lter band setting.
NARROW setting is used due to the interference from the heart beat
frequency, this interference has to be eliminated.
Program time
Threshold(µV)
Filter
Biofeedback
Work/Rest time(S)
Threshold setting
Max:99mins
0.6-2000μV, During work period the patient is
prompted to contract above the Threshold. In the
rest phase the patient is prompted to relax their
muscle.
Wide/Narrow
Above/Below/Continue/OFF:
Above the threshold, Below the threshold,
Continue-sound, OFF-no bar graph sound
Auto/Manual
Trial Number of work/rest repetitions, 2-99
2-99 sec
EMG Parameter
Custom Program 1
T