ADVENTURER+ Service Manual P/N 83020790 A
Ohaus Corporation www.ohaus.com 18
Figure 4-1 Segment Display
4.2.2 Calibration
To verify that the Balance is functioning properly, it must be calibrated prior to testing. At this point, the Balance
should have the menus set up if functional. After calibration, proceed to Performance Tests, paragraph 4.2.3 for
Precision balances and paragraph 4.2.4 for Analytical balances.
Calibration can be affected by changes in location, temperature, or rough handling. Adventurer
TM
balances can
be calibrated in two ways: Span calibration or Linearity calibration. Span calibration resets the balance's weighing
range using two weight values: zero and a weight value at or near the balance's capacity. Linearity calibration
minimizes deviation between actual and displayed weights within the balance's weighing range. Three weight
values are used: zero, a weight value within the balance's weighing range and a weight value at or near the
balance's specified capacity.
4.2.2.1 Calibration Masses
Before beginning calibration, make sure masses are available. If you begin calibration and realize calibration
masses are not available, exit the menu. The balance will retain previously stored calibration data. Calibration
should be performed as necessary to ensure accurate weighing. Masses required to perform the procedures are
listed in Table 4-1.
TABLE 4-1 CALIBRATION MASSES
Model Capacity Span Masses Linearity Masses
AR64 65 50 25/50
AR124 120 100 50/100
AR224 220 200 100/200
AR153 151 150 75/150
AR223 220 200 100/200
AR323 320 300 150/300
AR423 420 400 200/400
100 -- --
AR423D
420 400 200/400
AR522 520 500 250/500
AR1502 1520 1500 750/1500
AR2202 2200 2000 1000/2000
AR3202 3200 3000 1500/3000
AR4202 4200 4000 2000/4000
1000 -- --
AR4202D
4200 4000 2000/4000
AR4201CN 4200 4000 2000/4000
OHAUS adventurer+ - 4.2.2 CALIBRATION